    • Enclosed Medical Switching Power Supply
    • Universal Input Range 90-264VAC, 47-63 Hz
    • Cooling by Free Air Convection
    • 85 Watts and 120 Watt with 10CFM Forced Air
    • Ultra Compact Size: 3.15 x 2.35 x 1.7 Inches
    • 4000VAC Input to Output 2MOPP Insulation
    • High Efficiency up to 91%
    • With P.F.C. Function >0.9
    • <0.3W No Load Input Power
    • EMI for Both Class I (with PE) and Class II (without PE) Configuration
    • Suitable for BF Application with Appropriate System Consideration
    • UL / IEC / EN 60601 3.1 Edition Safety Approvals
    • 3-Year Product Warranty
    • All specifications valid at nominal input voltage,full load and +25°C after warm-up time unless otherwise stated.
    • Measured with 0.1U 50V // 47U 50V Ceramic Cap. Cross to output.
    • Please refer to the PDF file published on the web site as the final version.
    • We reserve the right to make alterations in the product materials and specifications without prior notification and consent to improve reliability, function or design or otherwise.
    • Display & Signage
    • Household & Smart Buildings
    • Industry 4.0
    • Industrial Automation
    • Lighting Control
    • Medical and Biotechnology
    • Security
    • Telecom
    • Transportation
  • Input Specifications

    Voltage: 90~264 VAC

    Frequency 47-63 Hz

    Inrush Current (<2ms) <40 A (115 VAC) / <70 A (230 VAC) (max.)

    Leakage Current <0.1 mA / 264 VAC (Touch Current) (max.)

    Output Specifications

    Power 120 (max.)

    Voltage 12, 24, 48

    Voltage Adj Range ±10% Output Voltage

    Voltage Accuracy (Full Load and Nominal Vin) ±2%

    Minimum Load (at Full Load) 0%

    Line Regulation (LL to HL at Full Load) ±1% (typ.)

    Load Regulation ±1% (typ.)

    Ripple & Noise 12S:160mV (Vp-p) ; others:1% Vout (max)

    Protection

    Over Power Protection Auto recovery, Hiccup mode

    Over Voltage Protection Latch off

    Over Temperature Protection Latch off

    Short Circuit Protection Protection level 1 (nominal)? Continuous, Auto recovery Protection level 2 (instantaneous high current)? Latch

    Environment Specification

    Operating Temperature -30 °C to +70 °C

    Storage Temperature -30 °C to +85 °C

    Temperature Coefficient ±0.05% / °C

    Altitude During Operation 5000m

    Humidity: 95% RH

    Atmospheric Pressure 56 kPa to 106 kPa

    Vibration IEC60068-2-6 (10~500Hz, 2G 10min./1cycle,60min. each along X, Y, Z axes)

    Shock IEC60068-2-27

    Safety and EMC

    Isolation Voltage (Input to Output) 4000VAC or 5656VDC

    Agency Approvals UL / IEC / EN 60601 3.1rd Edition (2 x MOPP)

    EMI (Conducted & Radiated Emission) Conducted: EN 55011 Conducted Class B Radiated: EN 55011 Class I class B / Class II class A

    EMS (Noise Immunity) EN 60601-1-2 4th Edition

    Others

    Dimensions (Tolerance ±0.5mm) 80.0 x 59.7 x 43.2 mm 3.15 x 2.35 x 1.7 inches

    Weight 258 g

    Case Material Aluminum

    Cooling Method Free convection / 10 CFM FAN

  • Datasheet

    UL 60601-1_MQCS100_MQF120

    IEC 60601-1_MQCS100_MQF120

    RoHS _ SGS-Switching Power Supply

    REACH