Skip to content

MQF500E Enclosed Medical Switching PSU 500 Watt

SKU MQF500E
by Arch
    • Enclosed Medical Switching Power Supply
    • Universal Input Range 90-264VAC, 47-63 Hz
    • 200 Watt with Free Air Convection
    • 500 Watt with 30CFM FAN
    • Ultra Compact Size: 5.5 x 3.25 x 2.42 Inches
    • Remote ON/OFF Function
    • Standby 5V@1A
    • High Efficiency up to 92%
    • With P.F.C. Function >0.94
    • Current Share Function for Option (except for 15S)
    • Suitable for BF Application with Appropriate System Consideration
    • 3-Year Product Warranty
    • All specifications valid at nominal input voltage,full load and +25°C after warm-up time unless otherwise stated.
    • Measured with 0.1U 50V // 47U 50V Ceramic Cap. Cross to output.
    • Please refer to the PDF file published on the web site as the final version.
    • We reserve the right to make alterations in the product materials and specifications without prior notification and consent to improve reliability, function or design or otherwise.
    • Display & Signage
    • Household & Smart Buildings
    • Industry 4.0
    • Industrial Automation
    • Lighting Control
    • Medical and Biotechnology
    • Security
    • Telecom
    • Transportation
  • Input Specifications

    Voltage: 90~264 VAC or 127~370 VDC

    Frequency 47-63 Hz

    Inrush Current (<2ms,Cold Start) <40 A (115 VAC) / <80 A (230 VAC) (max.)

    Leakage Current <0.1 mA / 264 VAC (Touch Current) (max.)

    Output Specifications

    Power 500 (max.)

    Voltage 12, 15, 24, 48

    Voltage Adj Range ±4% Output Voltage

    Voltage Accuracy (Full Load and Nominal Vin) ±2%

    Minimum Load (at Full Load) 3%

    Line Regulation (115-264 VAC) ±0.5% (typ.)

    Load Regulation (10% to 100% at Full Load) ±1% (typ)

    Ripple & Noise 12S,15S:160mV (Vp-p) ; 24S:240mV (Vp-p) ; 48S:480mV (Vp-p) (max)

    Protection

    Over Power Protection Auto recovery

    Over Voltage Protection Auto recovery

    Over Temperature Protection Auto Recovery

    Short Circuit Protection Protection level 1 (nominal)? Continuous, Auto recovery Protection level 2 (instantaneous high current)? Latch

    Environment Specification

    Operating Temperature -30 °C to +70 °C

    Storage Temperature -30 °C to +85 °C

    Temperature Coefficient ±0.03% / °C ( 0 °C~50 °C ) ±0.06% / °C ( -30 °C~0 °C )

    Altitude During Operation 5000m

    Humidity: 95% RH

    Atmospheric Pressure 56 kPa to 106 kPa

    Vibration IEC60068-2-6 (10~500Hz, 2G 10min./1cycle,60min. each along X, Y, Z axes)

    Shock IEC60068-2-27

    Safety and EMC

    Isolation Voltage (Input to Output) 4000VAC or 5656VDC

    Agency Approvals 12S/24S/48S: UL / IEC / EN 60601 3.1rd Edition (2 x MOPP), UL / IEC / EN 60950 AM2, UL/ IEC / EN 62368 ; 15S: UL / IEC / EN 60601 3.1rd Edition (2 x MOPP), UL / IEC / EN 60950 AM2 (meet), UL/ IEC / EN 62368 (meet)

    EMI (Conducted & Radiated Emission) Conducted: EN 55011 Conducted Class B Radiated: EN 55011 Class A

    EMS (Noise Immunity) EN 60601-1-2 4th Edition

    Others

    Dimensions (Tolerance ±0.5mm) 139.7 x 82.55 x 61.4 mm 5.5 x 3.25 x 2.42 inches

    Weight 690 g

    Case Material Aluminum

    Cooling Method Free convection / 30 CFM FAN

  • Datasheet

    MQF500 Current Sharing Application Note

    CE_Approval_EN 55032,24_MQF500

    EN 55011,60601-1-1-2_MQF500

    RoHS _ SGS-Switching Power Supply

    REACH

    REACH-SVHC 201 _ XVII 70 _ POPs