MQF500E Enclosed Medical Switching PSU 500 Watt
-
- Enclosed Medical Switching Power Supply
- Universal Input Range 90-264VAC, 47-63 Hz
- 200 Watt with Free Air Convection
- 500 Watt with 30CFM FAN
- Ultra Compact Size: 5.5 x 3.25 x 2.42 Inches
- Remote ON/OFF Function
- Standby 5V@1A
- High Efficiency up to 92%
- With P.F.C. Function >0.94
- Current Share Function for Option (except for 15S)
- Suitable for BF Application with Appropriate System Consideration
- 3-Year Product Warranty
-
- All specifications valid at nominal input voltage,full load and +25°C after warm-up time unless otherwise stated.
- Measured with 0.1U 50V // 47U 50V Ceramic Cap. Cross to output.
- Please refer to the PDF file published on the web site as the final version.
- We reserve the right to make alterations in the product materials and specifications without prior notification and consent to improve reliability, function or design or otherwise.
-
- Display & Signage
- Household & Smart Buildings
- Industry 4.0
- Industrial Automation
- Lighting Control
- Medical and Biotechnology
- Security
- Telecom
- Transportation
-
Input Specifications
Voltage: 90~264 VAC or 127~370 VDC
Frequency 47-63 Hz
Inrush Current (<2ms,Cold Start) <40 A (115 VAC) / <80 A (230 VAC) (max.)
Leakage Current <0.1 mA / 264 VAC (Touch Current) (max.)
Output Specifications
Power 500 (max.)
Voltage 12, 15, 24, 48
Voltage Adj Range ±4% Output Voltage
Voltage Accuracy (Full Load and Nominal Vin) ±2%
Minimum Load (at Full Load) 3%
Line Regulation (115-264 VAC) ±0.5% (typ.)
Load Regulation (10% to 100% at Full Load) ±1% (typ)
Ripple & Noise 12S,15S:160mV (Vp-p) ; 24S:240mV (Vp-p) ; 48S:480mV (Vp-p) (max)
Protection
Over Power Protection Auto recovery
Over Voltage Protection Auto recovery
Over Temperature Protection Auto Recovery
Short Circuit Protection Protection level 1 (nominal)? Continuous, Auto recovery Protection level 2 (instantaneous high current)? Latch
Environment Specification
Operating Temperature -30 °C to +70 °C
Storage Temperature -30 °C to +85 °C
Temperature Coefficient ±0.03% / °C ( 0 °C~50 °C ) ±0.06% / °C ( -30 °C~0 °C )
Altitude During Operation 5000m
Humidity: 95% RH
Atmospheric Pressure 56 kPa to 106 kPa
Vibration IEC60068-2-6 (10~500Hz, 2G 10min./1cycle,60min. each along X, Y, Z axes)
Shock IEC60068-2-27
Safety and EMC
Isolation Voltage (Input to Output) 4000VAC or 5656VDC
Agency Approvals 12S/24S/48S: UL / IEC / EN 60601 3.1rd Edition (2 x MOPP), UL / IEC / EN 60950 AM2, UL/ IEC / EN 62368 ; 15S: UL / IEC / EN 60601 3.1rd Edition (2 x MOPP), UL / IEC / EN 60950 AM2 (meet), UL/ IEC / EN 62368 (meet)
EMI (Conducted & Radiated Emission) Conducted: EN 55011 Conducted Class B Radiated: EN 55011 Class A
EMS (Noise Immunity) EN 60601-1-2 4th Edition
Others
Dimensions (Tolerance ±0.5mm) 139.7 x 82.55 x 61.4 mm 5.5 x 3.25 x 2.42 inches
Weight 690 g
Case Material Aluminum
Cooling Method Free convection / 30 CFM FAN
-
MQF500 Current Sharing Application Note
CE_Approval_EN 55032,24_MQF500