MQF240E
EU CUSTOMERS PLEASE NOTE:
Import duties, taxes and any additional delivery or clearance charges, are not included in the item price or the quoted shipping cost. These charges are the responsibility of the Buyer.
Please check with your local customs office to determine any additional charges prior to purchase.
Input Specifications
Voltage: 90~264 VAC
Frequency 47-63 Hz
Inrush Current (<2ms) <45 A (115 VAC) / <90 A (230 VAC) (max.)
Leakage Current <0.1 mA / 264 VAC (Touch Current) (max.)
Output Specifications
Power 240 (max.)
Voltage 12, 15, 24, 48
Voltage Adj Range ±4% Output Voltage
Voltage Accuracy (Full Load and Nominal Vin) ±2%
Minimum Load (at Full Load) 0%
Line Regulation ±1% (typ.)
Load Regulation (0% to 100% at Full Load) ±1% (typ)
Ripple & Noise 1% Vout (max.)
Protection
Over Power Protection Auto recovery, Hiccup mode
Over Voltage Protection Auto recovery
Over Temperature Protection Auto Recovery
Short Circuit Protection Protection level 1 (nominal)? Continuous, Auto recovery Protection level 2 (instantaneous high current)? Latch
Environment Specification
Operating Temperature -30 °C to +70 °C
Storage Temperature -30 °C to +85 °C
Temperature Coefficient ±0.05% / °C
Altitude During Operation 5000m
Humidity: 95% RH
Atmospheric Pressure 56 kPa to 106 kPa
Vibration IEC60068-2-6 (10~500Hz, 2G 10min./1cycle,60min. each along X, Y, Z axes)
Shock IEC60068-2-27
Safety and EMC
Isolation Voltage (Input to Output) 4000VAC or 5656VDC
Agency Approvals 12S/24S/48S: UL / IEC / EN 60601 3.1rd Edition (2 x MOPP), UL / IEC / EN 60950 AM2, UL/ IEC / EN 62368 ; 15S: UL / IEC / EN 60601 3.1rd Edition (2 x MOPP), UL / IEC / EN 60950 AM2 (meet), UL/ IEC / EN 62368 (meet)
EMI (Conducted & Radiated Emission) Conducted: EN 55011 Conducted Class B Radiated: EN 55011 Class I class B / Class II class A
EMS (Noise Immunity) EN 60601-1-2 4th Edition
Others
Dimensions (Tolerance ±0.5mm) 104.0 x 62.5 x 39.26 mm 4.1 x 2.46 x 1.544 inches
Weight 297 g
Case Material Aluminum
Cooling Method Free convection / 10 CFM FAN